Stability of Selected Biochemical Analytes in Serum Stored at Room Temperature for 24 Hours at Kericho County Referral Hospital, Kenya

  • Erick Kiprotich Ngetich Masinde Muliro University of Science and Technology
  • Sabella Kiprono, PhD Masinde Muliro University of Science and Technology
  • Dominic Marera, PhD Masinde Muliro University of Science and Technology
Keywords: Biochemical, Analytes, Stability, Room Temperature, Reference Change Value, Mean Percentage Difference
Share Article:

Abstract

Accurate and reliable medical laboratory test results are essential for correct patient diagnosis and management. This is however influenced by many preanalytical, analytical, and post-analytical factors. A general problem faced by most clinical laboratories is the inability to timely process all blood samples received due to overwhelming workload, reagents stock-outs, and equipment breakdown, necessitating storage which predisposes analytes to deterioration. To ensure accurate results from stored samples, the determination of the stability of the analytes under various environmental conditions and storage periods is important in informing acceptable sample handling and storage. This study was designed to experimentally observe changes and evaluate the stability of eleven biochemical analytes in serum samples from 20 randomly sampled healthy volunteers stored at room temperature over 24 hours at Kericho County Referral Hospital laboratory. 10mls of venous blood was drawn from each volunteer, serum was separated and five aliquot sets were prepared. Analysis was done on a set immediately, at 2, 4, 8, and 24 hours, to determine the concentration of the selected analytes on Human Star 100 chemistry analyser. Student pair T-test and Wilcoxon rank were used to analyse the data. Chloride, Creatinine, Glucose, Potassium, Sodium, Urea, Alkaline Phosphatase, Alanine Amino Transferase and Aspartate Amino Transferase registered stability up to 24 hours, with mean percentage differences less the respective calculated reference change value of 4.54, 35.97, 13.14, 10.89, 2.16, 32.15, 16.48, 42.25 and 42.52. Direct and total bilirubin showed stability only up to 2 and 4 hours, respectively, with mean percentage differences of -2.6 and 1.07 against respective calculated reference change values of 1.14 and 0.675 in the subsequent analysis. This study recommends immediate analysis of serum for total and direct bilirubin, while electrolytes, enzymatic analytes and other metabolic waste and substrates can be analysed within 24 hours without extra storage precautions. If testing is delayed, then effective storage including a dark room, refrigeration, or freezing, should be applied to maintain analytes stability for reliable results.

Downloads

Download data is not yet available.

References

Aarsand A. K., Røraas T., Fernandez-Calle P., Ricos C., Díaz-Garzón J., Jonker N., Perich C., González-Lao E., Carobene A, Minchinela J., Coşkun A., Simón M., Álvarez V., Bartlett W. A., Fernández-Fernández P., Boned B., Braga F., Corte Z., Aslan B., … & Sandberg S. (2018). The Biological Variation Data Critical Appraisal Checklist: A Standard for Evaluating Studies on Biological Variation. Clin Chem, 64:501-514

Alavi, N., Khan, S. H., Saadia, A., & Naeem, T. (2020). Challenges in preanalytical phase of laboratory medicine: rate of blood sample nonconformity in a tertiary care hospital. EJIFCC, 31(1), 21.

Asmelash, D., Worede, A., & Teshome, M. (2020). Extra-analytical clinical laboratory errors in Africa: a systematic review and meta-analysis. EJIFCC, 31(3), 208.

Beale, D. J., Jones, O. A., Karpe, A. V., Dayalan, S., Oh, D. Y., Kouremenos, K. A., ... & Palombo, E. A. (2016). A review of analytical techniques and their application in disease diagnosis in breathomics and salivaomics research. International journal of molecular sciences, 18(1), 24.

Bobby, L., Boyanton, Jr., & Blick, K. (2002). Stability of twenty-four analytes in human serum. Clinical Chemistry, 48(12), 2242-2247.

Bonini, P., Plebani, M., Ceriotti, F., & Rubboli, F. (2002). Errors in Laboratory Medicine. Clinical Chemistry, 48(5), 691 - 698.

Camargo, Fabricio et al. 2020. “Serum Biochemistry of the Arrau Turtle (Podocnemis Expansa): Frozen Analyte Stability and the Effects of Long-Term Storage.” Veterinary Clinical Pathology 49(1): 42–47.

Carey, R. N., Jani, C., Johnson, C., Pearce, J., Hui-Ng, P., & Lacson, E. (2016). Chemistry testing on plasma versus serum samples in dialysis patients: clinical and quality improvement implications. Clinical Journal of the American Society of Nephrology, 11(9), 1675-1679.

Clinical and Laboratory Standards Institute (CLSI). (2010). Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline. Wayne PA, 2010.

Damtie, T. A., Ibrahim, A. J., & Yikna, B. B. (2020). Supply chain management performance of HIV/AIDS commodities and factors affecting it at health facilities of SNNPRS of Ethiopia; from the perspective of achieving 90-90-90 strategies. Integrated Pharmacy Research & Practice, 9, 11.

Dittadi, R., Fabricio, A. S., Rainato, G., Peroni, E., Di Tonno, F., Vezzù, B., ... & Gion, M. (2019). Preanalytical stability of [-2] proPSA in whole blood stored at room temperature before separation of serum and plasma: implications to Phi determination. Clinical Chemistry and Laboratory Medicine (CCLM), 57(4), 521-531.

Emre, H. O., Karpuzoglu, F. H., Coskun, C., Sezer, E. D., Ozturk, O. G., Ucar, F., ... & Coskun, A. (2021). Utilisation of biological variation data in the interpretation of laboratory test results–survey about clinicians’ opinion and knowledge. Biochemia medica, 31(1), 93-102.

Fisher, W. E., Cruz-Monserrate, Z., McElhany, A. L., Lesinski, G. B., Hart, P. A., Ghos, R., ... & Conwell, D. L. (2018). Standard operating procedures for biospecimen collection, processing, and storage: from the consortium for the study of chronic pancreatitis, diabetes, and pancreatic cancer. Pancreas, 47(10), 1213.

Flores, C. F. Y., Pineda, Á. D. L. M. H., Bonilla, V. M. C., & Sáenz-Flor, K. (2020). Sample management: stability of plasma and serum on different storage conditions. EJIFCC, 31(1), 46.

Gómez-Rioja, R., Amaro, M. S., Diaz-Garzón, J., Bauçà, J. M., Espartosa, D. M., & Fernández-Calle, P. (2019). A protocol for testing the stability of biochemical analytes. Technical document. Clinical Chemistry and Laboratory Medicine (CCLM), 57(12), 1829-1836.

Hedayati, M., Razavi, S. A., Boroomand, S., & Kheradmand Kia, S. (2020). The impact of preanalytical variations on biochemical analytes stability: A systematic review. Journal of Clinical Laboratory Analysis, 34(12), e23551.

Ikeda, K., Ichihara, K., Hashiguchi, T., Hidaka, Y., Kang, D., Maekawa, M., ... & Committee for Standardization, The Japanese Society of Laboratory Medicine (JSLM). (2015). Evaluation of the short-term stability of specimens for clinical laboratory testing. Biopreservation and Biobanking, 13(2), 135-143.

Kachhawa, K., Kachhawa, P., Varma, M., Behera, R., Agrawal, D., & Kumar, S. (2017). Study of the stability of various biochemical analytes in samples stored at different predefined storage conditions at an accredited laboratory of India. Journal of laboratory physicians, 9(01), 011-015.

Magnette, A., Chatelain, M., Chatelain, B., Ten Cate, H., & Mullier, F. (2016). Preanalytical issues in the haemostasis laboratory: guidance for the clinical laboratories. Thrombosis journal, 14(1), 1-14.

Mrazek, C., Lippi, G., Keppel, M. H., Felder, T. K., Oberkofler, H., Haschke-Becher, E., & Cadamuro, J. (2020). Errors within the total laboratory testing process, from test selection to medical decision-making–A review of causes, consequences, surveillance and solutions. Biochemia Medica, 30(2), 215-233.

Najat, D. (2017). Prevalence of preanalytical errors in clinical chemistry diagnostic labs in Sulaimani city of Iraqi Kurdistan. PloS one, 12(1), e0170211.

Nwosu, O. K., Aloh, G. S., & Ihedioha, J. L. (2009). Changes in alt, ast and alp values of plasma and serum samples stored at refrigerator (4 0c) and room temperature (32 0c) for up to five days. Bio-Research, 7(2).

Razi, F., Bandarian, F., Niazpour, F., Shafiee, G., Heshmat, R., Abdollahi, M., ... & Nasli-Esfahani, E. (2020). Stability of Thirty-Four Analytes in Blood Samples of Diabetic Patients. Clinical Laboratory, 66(1).

Shimizu, Y., & Ichihara, K. (2019). Elucidation of stability profiles of common chemistry analytes in serum stored at six graded temperatures. Clinical Chemistry and Laboratory Medicine (CCLM), 57(9), 1388-1396.

Sofronescu, A. G., Loebs, T., & Zhu, Y. (2012). Effects of temperature and light on the stability of bilirubin in plasma samples. Clinica chimica acta; international journal of clinical chemistry, 413(3-4), 463–466. https://doi.org/10.1016/j.cca.2011.10.036.

Yamane, T. (1967). Statistics: An Introductory Analysis, 2nd Ed., New York: Harper and Row.

Zemlin, A. E. (2018). Errors in the extra-analytical phases of clinical chemistry laboratory testing. Indian Journal of Clinical Biochemistry, 33(2), 154-162.

Published
19 September, 2022
How to Cite